Please use this identifier to cite or link to this item:
http://hdl.handle.net/123456789/1585
Title: | Assessment of the accuracy of analyzers for automatic determination of PM10 and PM2.5 Particulate Matter in ambient air | Authors: | Bucur, Elena Motisan, Radu Vasile, Andrei Tanase, Gheorghita Pascu, Luoana Florentina Lehr, Carol Blaziu Cuciureanu, Adriana |
Keywords: | Accuracy;Ambient air;Pearson correlation;PM2.5;PM10 | Issue Date: | Jan-2020 | Publisher: | SYSCOM 18 SRL | Abstract: | The paper presents the test results regarding the evaluation of the accuracy of the PM2.5 and PM10 particulate matter concentration measurement performed with the uRADMonitor A3 fixed air quality monitoring station produced by SC MAGNASCI SRL. The procedure involves the calculation of the accuracy elements: trueness and precision, based on the experimental data obtained by measuring the concentration of particulate matter using the tested analysers in parallel with the reference method, SR EN 12341: 2014, and analysis of data series by Pearson correlation and linear regression. |
Description: | Revista de Chimie, vol. 71, no. 1, 2020, pp. 83-87 |
URI: | http://hdl.handle.net/123456789/1585 | ISSN: | 1582-9049 |
Appears in Collections: | Articles |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
13 BUCUR E 1 20.pdf | 396.74 kB | Adobe PDF | View/Open |
Google ScholarTM