Please use this identifier to cite or link to this item: http://hdl.handle.net/123456789/1585
Title: Assessment of the accuracy of analyzers for automatic determination of PM10 and PM2.5 Particulate Matter in ambient air
Authors: Bucur, Elena
Motisan, Radu
Vasile, Andrei
Tanase, Gheorghita
Pascu, Luoana Florentina
Lehr, Carol Blaziu
Cuciureanu, Adriana
Keywords: Accuracy
Ambient air
Pearson correlation
PM2.5
PM10
Issue Date: Jan-2020
Publisher: SYSCOM 18 SRL
Abstract: The paper presents the test results regarding the evaluation of the accuracy of the PM2.5 and PM10 particulate matter concentration measurement performed with the uRADMonitor A3 fixed air quality monitoring station produced by SC MAGNASCI SRL. The procedure involves the calculation of the accuracy elements: trueness and precision, based on the experimental data obtained by measuring the concentration of particulate matter using the tested analysers in parallel with the reference method, SR EN 12341: 2014, and analysis of data series by Pearson correlation and linear regression.
Description: Revista de Chimie, vol. 71, no. 1, 2020, pp. 83-87
URI: http://hdl.handle.net/123456789/1585
ISSN: 1582-9049
Appears in Collections:Articles

Files in This Item:
File Description SizeFormat 
13 BUCUR E 1 20.pdf396.74 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.