Title: ICP-EOS method development and validation for Cr(VI) determination in plants
Authors: Tenea, Anda Gabriela 
Mureseanu, Mihaela 
Vasile, Gabriela 
Dinu, Cristina 
Affiliations: National Research and Development Institute for Industrial Ecology, ECOIND 
University of Craiova 
National Research and Development Institute for Industrial Ecology, ECOIND 
National Research and Development Institute for Industrial Ecology, ECOIND 
Keywords: Cr(VI);ICP-EOS;Mint;Basil;Speciation
Issue Date: 27-Dec-2021
Publisher: University of Craiova
Abstract: 
The aim of this study was the development and validation of a new inductively coupled plasma optical emission spectrometry (ICP-EOS) method for the determination of Cr(VI) from vegetable samples. The new method was validated entirely by verifying performance parameters (linearity, inter-day and intra-day precision,
limit of detection - LOD, and limit of quantification - LOQ). The vegetable samples, mint, and basil were contaminated with a known amount of K2Cr2O7. Cr(VI) was extracted from the vegetable samples by ultrasound-assisted extraction technique, using 0.1 M Na2CO3 as the extraction medium. The ICP-EOS method proved very sensitive, allowing us to determine up to 3.0 µg/g Cr(VI) in vegetable matrices, which was established as the method LOQ.
Description: 
Annals of the University of Craiova, The Chemistry Series, vol. XLVII, no. 2, 2021, pp. 34-44
URI: http://hdl.handle.net/123456789/1884
Appears in Collections:Articles

Files in This Item:
File Description SizeFormat
Lucrare Anda Tenea 2021 Anale Craiova.pdfArticle608.41 kBAdobe PDFView/Open
Show full item record

Google ScholarTM

Check


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.