Title: Electrochemical System for Field Control of Hg2+ Concentration in Wastewater Samples
Authors: Tenea, Anda Gabriela 
Dinu, Cristina 
Buica, George Octavian 
Vasile, Gabriela 
Affiliations: National Research and Development Institute for Industrial Ecology, ECOIND 
National Research and Development Institute for Industrial Ecology, ECOIND 
University Politehnica of Bucharest, Romania 
National Research and Development Institute for Industrial Ecology, ECOIND 
Keywords: Carbon screen-printed modified electrode;Mercury;Electrochemical detection;Wastewater analysis
Issue Date: Jan-2023
Publisher: MDPI
Abstract: 
The paper presents the validation of an electrochemical procedure for on-site Hg2+
ions determination in wastewater samples using a modified carbon screen-printed electrode (SPE)
with a complexing polymeric film based on poly(2,2 0-(ethane-1,2-diylbis((2-(azulen-2-ylamino)-2-
oxoethyl)azanediyl))diacetic acid) (polyL). Using metal ions accumulation in an open circuit followed by anodic stripping voltammetry, the SPE-polyL electrode presents a linear range in the
range of 20 µg/L to 150 µg/L, with a limit of detection (LOD) = 6 µg/L, limit of quantification
(LOQ) = 20 µg/L, and average measurement uncertainty of 26% of mercury ions. The results obtained in situ and in the laboratory using the SPE-polyL modified electrode were compared with those
obtained by the atomic absorption spectrometry coupled with the cold vapor generation standardized
method, with the average values indicating excellent recovery yields.
Description: 
Sensors, 2023, 23(3), 2023, 1084; https://doi.org/10.3390/s23031084
URI: http://hdl.handle.net/123456789/1967
Appears in Collections:Articles

Files in This Item:
File Description SizeFormat
sensors-23-01084.pdfArticle841.32 kBAdobe PDFView/Open
Show full item record

Google ScholarTM

Check


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.