Options
New ion-chromatography method for detection of chlorite, chlorate, and bromate in drinking water
ISSN
2668-8530
Date issued
2020-10
Author(s)
DOI
10.21698/rjeec.2020.207
Abstract
A direct ion chromatographic method for determination of chlorite, chlorate and bromate in the presence of fluoride, chloride, nitrate, nitrite and bromide in treated drinking waters was described. Separation of target analytes was achieved using an AS19-HC analytical column (250 mm x 4 mm), AG 19-HC guard column (50 mm x 4 mm) and KOH 25 mmol/L as mobile phase. Inorganic analytes were eluted using a flow rate of 1 mL/min. Column temperature was set up and maintained at 300C. The analyte ions were quantified using a suppressed conductivity detector.
Files
Loading...
Name
07.pdf
Description
Proceedings paper
Size
632.52 KB
Format
Adobe PDF
Checksum
(MD5):da5611c252c7d5686058b0d1f3aad719